Fluorophos ALP Diagnostic Test – Substrate Test 1
1/ Pipette 2ml of substrate into a cuvette and warm.
2/ In A/D mode place cuvette into instrument.
If the result is greater than 600 and less than 1200 the substrate is within specification, click here to proceed to the Pipette test.
If the result is greater than than 1200 click here to proceed.
If the result is less than 600 click here to proceed.