Fluorophos ALP Diagnostic Test – Pipette Test 1

1/ Check the calibration of the 75µl sample pipette using the “calibration gauge” which is situated in the end of the pipette (see below).

2/ If the calibration is out re-calibrate the pipette. Adjustment is obtained by loosening the calibration set screw on the side of the pipette. Set the plunger depth using the calibration gauge then re-tighten the set-screw. Return the gauge to its storage position at the end of the pipette.

3/ Replace the pipette tip.

Fluorophos Pipette

If the calibration is successful the next step is to check the pipette, click here to continue.

If the calibration not successful please contact QCL.

QCL Scientific
Privacy Overview

This website uses cookies so that we can provide you with the best user experience possible. Cookie information is stored in your browser and performs functions such as recognising you when you return to our website and helping our team to understand which sections of the website you find most interesting and useful.