Fluorophos ALP Diagnostic Test – Pipette Test 1

1/ Check the calibration of the 75µl sample pipette using the “calibration gauge” which is situated in the end of the pipette (see below).

2/ If the calibration is out re-calibrate the pipette. Adjustment is obtained by loosening the calibration set screw on the side of the pipette. Set the plunger depth using the calibration gauge then re-tighten the set-screw. Return the gauge to its storage position at the end of the pipette.

3/ Replace the pipette tip.

Fluorophos Pipette

If the calibration is successful the next step is to check the pipette, click here to continue.

If the calibration not successful please contact QCL.