Fluorophos ALP Diagnostic Test – Instrument Test 4

1/ Place a cuvette with DI water into the cuvette well of the instrument and insert a thermometer.

If the temperature is 38oC (+/- 1) the instrument temperature is correct. You have now performed a series of tests to verify the instrument and heater block are functioning correctly. The next step is to check the substrate, click here to continue.

There is no adjustment possible to the instrument temperature. If the reading is outside this parameter please contact QCL.

QCL Scientific
Privacy Overview

This website uses cookies so that we can provide you with the best user experience possible. Cookie information is stored in your browser and performs functions such as recognising you when you return to our website and helping our team to understand which sections of the website you find most interesting and useful.