Fluorophos ALP Diagnostic Test – Instrument Test 2

1/ Pipette 2 ml FLA280 Daily Instrument Control into a cuvette and warm in the heater block.

2/ With the instrument still in A/D mode place cuvette in well.

If the reading is 602 (+/- 12) click here to continue.

If the reading is outside this parameter click here to continue.

QCL Scientific
Privacy Overview

This website uses cookies so that we can provide you with the best user experience possible. Cookie information is stored in your browser and performs functions such as recognising you when you return to our website and helping our team to understand which sections of the website you find most interesting and useful.