Fluorophos ALP Diagnostic Test – Instrument Test 2a

1/ Ensure the Fluorophos has been switched on for at least 2 hours to ensure complete electronic stabilization.

2/ If the reading is stable but out of range the reading can be adjusted using the potentiometer on the side of the instrument. Note: this facility is only available on instruments less than approximately 10 years old.

Fluorophos Potentiometer

If the reading is 602 (+/- 12) click here to continue.

If the reading remains unstable, or there is no adjustment possible on your Fluorophos please contact QCL.

QCL Scientific
Privacy Overview

This website uses cookies so that we can provide you with the best user experience possible. Cookie information is stored in your browser and performs functions such as recognising you when you return to our website and helping our team to understand which sections of the website you find most interesting and useful.