Fluorophos ALP Diagnostic Test – Instrument Test 1

1/ Switch Instrument to A/D Mode (access the A/D test by pressing the setup key).

2/ Check the reading with empty cuvette well.

If the reading is 302 (+/- 4) click here to continue.

If the reading is outside this parameter click here to continue.