Pharmaceutical and Research Polarimeter

The Rudolph Research Analytical Autopol III is dual wavelength research grade instrument

Rudolph Research Analytical Autopol III<br />
Pharmaceutical and Research Polarimeter
Autopol III<br />
Pharmaceutical and Research Polarimeter

Polarimeter for pharmaceutical and research applications

589nm and 546nm wavelengths

High quality polarizing prisms

Temperature probe

Polarimeter for Pharmaceutical and Research Applications

The Autopol III is a two wavelength pharmaceutical and research grade digital automatic polarimeter. The Autopol III features dual wavelengths and accuracy similar to Rudolph’s more advanced models, but with a simplified interface. Just place the cell in the sample chamber, press the Start Key and get 5 results in less than 25 seconds.

Standard Features

  • Accuracy 0.002° Arc up to 1°, 0.2% above 1°, 0.01° Z(ISS)
  • 589nm and 546nm wavelengths
  • Complete accessory package with IQOQPQ documentation, 2 non-Temptrol sample cells, 1 Temptrol 100 mm cell, and 1 NIST traceable calibration standard with a NIST certificate
  • Built in thermoprobe for temperature measurement
  • Temperature probe is covered with a micron thick glass coating to stand up to high molar Hydrochloric Acids
  • Windows embedded for direct connection to the network server and flexible USB Printing

Optional Features

  • TempTrol automatic electronic heating and cooling 15° – 40°C ±0.2°
  • 405nm, 436nm, and 578nm, wavelengths
  • AP Accuracy Option: 0.002° up to 10° Arc
  • AP Resolution Option: 0.01, 0.001°, 0.0001° Arc Selectable

A cGLP/GMP compliant printout, which includes the temperature of the sample is generated by pressing the Print/Send Key after the measurement is finished.

Try Before You Buy
The Autopol III is available for a short term trial.

Request Information

2 + 1 =

Autopol III cGLP/GMP compliant printout,

Autopol III Automatic Polarimeter Accessory Packages
The Autopol III without Temptro lPeltier temperature control is supplied with one of three automatic polarimeter accessory packages Each package includes a NIST traceable calibration standard and certificate, two Tungsten Halogen lightsource lamps and an alignment target. Each has different cell configurations.

Polarimeter Accessory Package A
Includes one 100mm, 6ml glass centre fill cell and one 100mm, 6ml glass center fill stainless steel jacketed cell. Cell windows can be removed and cleaned to avoid sample residue build up. All cells in this package are compatible with acids.

Autopol III Polarimeter accessories 1

Polarimeter Accessory Package B
One 50mm glass jacketed fixed end cell and one 100mm glass center fill cell with removable windows. Fused end cells avoid improper assembly by untrained users, but are very fragile and difficult to clean. All cells in this package are compatible with acids.

Autopol III polarimeter accessories 2

Polarimeter Accessory Package C
Each cell has a lifetime warranty against breakage. Each is easily filled and cell windows are removable. Cells are replaceable and are available in 50mm, 100mm and 200mm length versions. Cell volumes range from 0.5ml to 15ml.

Autopol polarimeter III accessories 3

Also available the Autopol I and Autopol II

Download Literature

Specification

Measuring Mode: Optical Rotation Specific Rotation Specific Rotation Plus Concentration User Defined Scale Sugar Degrees °Z(ISS)
Measuring Scale Degrees: Arc Optical Rotation
Resolution: 0.001° Arc Optical Rotation 0.001% Concentration 0.001 Specific Rotation
Accuracy: 0.002° Arc up to 1° 0.2% above 1° 0.0 1° Z (ISS)
AP Acccuracy Option: Accuracy ±0.002° up to 10° ±0.003° 10° – 89 ° Arc Optical Rotation
AP Resolution Option: AP Selectable Resolution: 0.01, 0.001, 0.0001° Arc
Reproducibility: 0.002° Arc Optical Rotation
Measuring Range Arc Optical Rotation: ± 89.99°
Measuring Range Arc Specific Rotation: ± 999.99°
Measuring Range Concentration: 0-99.9%
Prism: Glan Thompson Calcite
Optical Wavelengths: 589nm 546nm Optional Wavelengths 365nm, 405nm, 436nm, 578nm.
Wavelength Selection: Touchscreen Selectable: 2 standard, 4 Optional
21 CFR Part 11 Compliant Software: With optional PC Software
Temperature Control: External water bath (Standard) TempTrol automatic electronic heating and cooling 15° – 40°C ±0.2° C (Optional)
Temperature Probe Measurement Range: 10° – 100°C
Temperature Probe Measurement Accuracy: ±0.1°C
Measurement Time: Average 5 measurements in less than 25 seconds
Light Source: Tungsten-halogen 6V, 20W, avg. 2,000 hour life
Optional Light Source: LED 20-30 mA 100,000 hr. avg. life
Sample Chamber: Accepts sample tubes up to 200mm
Data Storage: 32 GB Non-removable Compact Flash
Communication Interface: 3- USB Ports, 1-USB 3.0 2- RS232 serial ports, Ethernet Port for Network Communication
Calibration: Automatic calibration via touchscreen
Operating System: Windows Embedded
Display: 8 inch colour, 800 x 600 pixel resolution with 400 nits of brightness.
User Interface: Touchscreen
Automatic Sensitivity Control: Measures samples with transmittance as low as 0.01% (up to O.D. 4.0)
Calendar/Clock Functions: Battery-backed clock; time and date sent to computer & printer
Input Power: 100 – 240 V, 50/60 Hz
Dimensions: 617mm W x 323mm H x 432mm D
Weight: 19.05 kg

QCL Scientific
Privacy Overview

This website uses cookies so that we can provide you with the best user experience possible. Cookie information is stored in your browser and performs functions such as recognising you when you return to our website and helping our team to understand which sections of the website you find most interesting and useful.