Measurement Scales Refractive Index (nD), Brix (% Sucrose), and up to 100 custom programmed scales
Measurement Range Refractive Index 1.33000-1.53000, Brix 0-100, Urine SG 1.0000 - 1.0400
Resolution Refractive index ±0.00001, Brix 0.01, Urine SG 0.0001
Reproducibility Refractive index ±0.00002, Brix ±0.02
Accuracy Refractive index ±0.00002, Brix ±0.01
Optical Wavelength 589.3nm (NaD line)
Prism Synthetic Sapphire
Measurement Response Time User configurable, can be less than 30 seconds
Temperature Control Range 10°C to 50°C (within 10°C of ambient)
Temperature Control Reproducibility ±0.01°
Ambient Temperature Limit 5°C to 40°C
Temperature Correction Range 4°C to 95°C (for sucrose solutions)
Sample Temperature Limit -20°C to 250°C
Calibration Using water or NIST traceable fluids. Factory default calibration can always be reset.
Dimensions L 43.5cm, W 30.5cm, H 33cm, Weight 10.4kg
Power Requirements 100 - 240 volts, 50 Hz - 60 Hz