QCL
QCL QCL
 
Home
About Us
Contact
Ordering
Service
User Guides
Brochures
Newsletters
MSDS
Links
Terms & Conditions
 
Partner With QCL
Adobe Acrobat
Download Adobe
Acrobat reader
OMNIChem  

NEXTChem’s most advanced and capable unit. It can perform a great number of measurements on a large number of sample points. It can also provide extensive process control. OMNIChem is suitable for analyses in the following industries:

Wastewater Ammonia, Nitrate, Hydrogen Sulfide, Free/Total Alkalinity
Drinking Water Total Hardness, Calcium Hardness, Chlorine, Fluoride
Metal Pre-Treatment Free/Total Acid, Free/Total Alkalinity, Fluoride, Nitrite
Textile Indigo, Hydrogen Peroxide, Caustic
Mining Cyanide, Silver, Copper, Arsenic
Pulp/Paper ABC Analyses, Ozone, Hydrogen Peroxide
Petro-Chemical Refinery Caustic, Sulfide/Mercaptides, Hydrogen Sulfide
Chemical Acetic Acid, Phosphoric Acid, Caustic, Chloride
Power Generation Low Level Hardness, Boron, Sodium

Measurement Methods
Two types of measurement methods are used for almost all NEXTChem analyzers: titration and (ion selective) standard addition. Both types of measurements methods can be used on all NEXTChem analyzers making each type of analyzer extremely flexible.

Titration
Titration methods involve the addition of a reagent to the sample until an endpoint is detected with an electrode or color probe. Most of the titrations fall with one of five different categories: acid/base, oxidation/reduction, complexation, precipitation, or color.

Standard Addition
Measurements with the Ion Selective Electrodes (ISE) utilize the standard addition method. The standard addition method involves measuring the sample with the ISE before and after a know standard has been added and using the difference in the readings to determine concentration.

Precision Reagent Pump
The heart of a process analyzer is the way reagents are delivered to the measurement cell. With all NEXTChem analyzers precision reagent pumps are used to deliver reagents. Because of the use of advanced stepper motor technology these pumps are very accurate + 0.001 mL. The utilization of a superior seal material makes the precision reagent pump very low maintenance only requiring to be rebuilt once per year.

Ease of Programming
All NEXTChem analyzers use touchscreen which make them very user friendly. The analyzer is programmed in a similar manner to the interaction with a bank ATM. In addition, to intuitive operation of the touchscreen the user can select English or a second language: Spanish, Portuguese, French, Dutch, traditional Chinese, simplified Chinese, Arabic, and Korean. The non-English interface makes the analyzer very easy for non-English speakers to use.

Filter-Free Operations
The MINIChem Series like all NEXTChem analyzers are designed to operate FILTER-FREE; this type of sampling INCREASES RELIABILITY and DECEASES MAINTENANCE. To make FILTER-FREE operation possible NEXTChem has created the “self-cleaning valve”. Utilizing the “self-cleaning valve” particles up to <6 mm can be easily handled. Large bore tubing is used to transfer the sample from the sample panel to the measurement cell 1/4” (6 mm) or 3/8” (9mm). Combining the self-cleaning valve with large bore tubing virtual eliminates the possibility of plugging; usually the most important issue in the operation of a process analyzer.

Automated Sample Dilution
The MINIChem series of analyzers can be configured to measure very high concentrations. This is accomplished by utilizing a dilution measurement cell. The dilution measurement cell is used with an external sample capture loop (5-30mL). This allows sample with up to <2 mm particles to be captured FILTER-FREE and sample burettes are never used.

Process Control
OMNIChem and MINIChem Series Process Analyzers have the additional capability of on-board process control. This process control capability allows the analyzer to control chemical feed pumps being used to add chemistry to the chemical bath being controlled. This is done with pneumatic chemical feed pumps. The MINIChem can direct 1 chemical feed pump, the MINIChem-Plus 2 chemical feed pumps; the Basic OMNIChem can direct 4 chemical feed pumps, and the Advanced OMNIChem 8 chemical feed pumps.

Technical Specification
Non-Dilution Measurement Cell Sample Volume: 90 / 150 mL
Maximum Particle Size: <6mm
Dilution Measurement Cell Sample Volume: 5 – 30 mL
Maximum Particle Size: <2mm
Samples / Analyses Titration Dilution: + 0.5%, non-dilution + 0.2%
Ion Selective Dilution: + 5.0%, non-dilution + 2.0%
Analysis Time: 2 – 10 minutes, 5 minutes typical
Sample Points: 4 Basic, 8 Advanced
Sample Temp: 10 – 212 F (-15 – 100 C)
Total Analyses: 8 Basic, 12 Advanced
Reagent Pumps: 6 Basic, 6 Advanced
Process Control: 4 Basic, 8 Advanced
Utilities Power: 120/240 VAC, 60/50 Hz, 5/3 A
Compressed Air: 70 – 100 PSI (5.5 – 6.5 bar)
Sample Pressure: 5 – 50 PSI (0.3 – 3.5 bar)
Sample Flow: 1 – 10 GPM (4 – 40 LPM)
Water Pressure: 5 – 50 PSI (0.3 – 3.5 bar)
Water Flow: 1 – 10 GPM (4 – 40 LPM)
Enclosures NEMA: 4X (IP65) Fiberglass or NEMA 12 (IP 52) Steel
Fiberglass: 78” x 24” x 24” (200 cm x 61 cm x 61 cm)
Steel: 79” x 24” x 24” (200 cm x 61 cm x 61 cm)
Fiberglass: 200 lbs (90 Kg)
Steel: 300 lbs (135 Kg)
Touchscreen: 5.7” grayscale/color or 6.5” color
Temperature: 40 – 100 F (5 – 40 C)
Heaters/Coolers: Optional
Purge System: Optional
Inputs / Outputs Basic OMNIChem: 8 off, 4-20 mA (Outputs)
Advanced OMNIChem: 16 off, 4-20 mA (Outputs)
Alarm Relay: 1 off, 6 A maximum
Serial Outputs: 1 off, RS232 / Modbus
DeviceNet / Profibus-DP: 1 off, Optional (Slave Only)
Ethernet: 2 off, Optional (Data / Control)
Auxiliary Analog: 4 off, 4-20 mA (Inputs)
Certifications CE
CSA-us/c

ECOChem
Click image to enlarge

QCL