LTF Tube Furnaces: 1700°C 1800°C  

Models LTF 17 and LTF 18 - maximum operating temperatures 1700°C and 1800°C.
These furnaces are constructed using high grade insulation materials throughout for rapid response rates and maximum thermal efficiency and stability. Heating is provided by vertically suspended molybdenum disilicide elements.

Model LTF 175 - maximum operating temperature 1750°C.
This model is available in vertical format only. Heating is via elements which are mounted parallel to the tube axis. The main constituent of the elements is lanthanum chromite, with other inclusions being added to provide them with their unique properties. Lanthanum chromite is naturally inert and does not form a crystalline layer on the element when operated in air or pure oxygen. These elements exhibit little or no ageing and their electrical characteristics remain constant throughout their life. In the LTF multi-element furnace, the replacement of a single element does not affect the behaviour.

Technical Specification
  LTF 17/75/300 LTF 17/75/600 LTF 175/50/200
Maximum operating temperature (°C) 1700 1700 1750
Maximum continuous temperature (°C) 1650 1650 1700
Tube inside diameter (mm) 75 75 50
Heated length (mm) 300 600 200
Maximum power (kW) 5.5 9.0 8.3
External dimensions including console
(mm – h x w x d)
1550 x 720 x 615 1550 x 1020 x 615 860 x 660 x 660
Optional separate work tubes - Short
(mm – id x length)
75 x 700 75 x 1200 50 x 850
Optional separate work tubes - Long
(mm – id x length)
75 x 1000 75 x 1000 50 x 1220
       
  LTF 175/75/350 LTF 175/125/350 LTF 18/75/300
Maximum operating temperature (°C) 1750 1750 1800
Maximum continuous temperature (°C) 1700 1700 1750
Tube inside diameter (mm) 75 125 75
Heated length (mm) 350 350 300
Maximum power (kW) 12.0 12.0  
External dimensions including console (mm – h x w x d) 1010 x 660 x 660 1010 x 660 x 660 1550 x 720 x 615
Optional separate work tubes - Short
(mm – id x length)
75 x 1000 125 x 1000 75 x 900
Optional separate work tubes - Long
(mm – id x length)
75 x 1350 125 x 1500 75 x 1200