Dissolved Oxygen Meters
 
Clearing Errors in Orion Star Series Meters

Clearing out an Error 38
1/ Make sure you are in the Measure Mode

2/ Press the Set Up key

3/ Press the Down Arrow 5-6 times until you see L09 (this stands for LOG)

4/ Press the Line Select (3 bar) key twice to move the indicator arrow on the left side of the screen to the bottom line

5/ The reading should be YES and flashing. Press the Down Arrow key to change that to read NO

6/ Press the Line Select (3 bar) key once more to move the indicator arrow back to the top line

7/ Press the Measure key to accept the changes you just made, and proceed back into measure mode

Clearing out an Error 200
1/ In the Measure Mode, press Measure to clear the error.

2/ Press Set Up key

3/ With the arrow pointing at the top line press the Down Arrow one time

4/ The display will read AUTO on the top line and SAPL on the middle line

5/ Press the Line Select key two times to move the arrow to the bottom line

6/ Press the Down Arrow until it says OFF

7/ Press the Line Select key to move the arrow to the top line 8. Press Measure to exit out of setup