QCL
QCL QCL
 
Home
About Us
Contact
Ordering
Service
User Guides
Brochures
Newsletters
MSDS
Links
Terms & Conditions
 
Partner With QCL
Adobe Acrobat
Download Adobe
Acrobat reader
4250 Single Sample Cryoscope  

The 4250 single-sample cryoscope provides fast, accurate measurement of extraneous water in milk. Combining advanced features with simplified operation, this low maintenance, reliable high performance laboratory instrument supersedes the 4D3 and Fiske Mark 2 models. Up to 35 samples per hour can be measured and results can be displayed as desired and results can be collected on a computer or connected to a LIMS system.

The 4250 uses the industry standard thermistor freezing point method to precisely determine the level of extraneous water. With its on-board sample identification, broad measurement range and optimal sample size requirement, the 4250 provides accurate results compliant with ISO5764 plateau-seeking mode.

Quality management is improved by the 4250's unique key-switch, which allows the advanced set-up features and calibration modes to be locked out. In Operator Mode, routine testing can be accomplished with just two buttons, in Supervisor Mode, all operations are accessible.

The calibration procedure is quick and computer controlled. It will also detect and warn the operator if the calibration solutions are tested out of order. Settings are stored in a battery-protected memory. The 4250 also incorporates a new thermal printer.

Technical Specification
Sample Volume 2 or 2.5 mL
Test Time Approximately 90 seconds in the 30-second timed mode
Sample Capacity Single Sample
Units -m°C or -m°H
Resolution 1 m°C or m°H
Range 0 to 1000 m°C or m°H
Communications On-board printer, DTE RS-232 serial port, and optional barcode scanner
Performance at Reference Conditions*

- 20 to 25°C (68 to 77°F); 40 to 60% relative humidity; tolerances of reference or calibration solutions excluded:
Linearity - Less than ±0.5% from a straight line
Repeatability - ±2 m°C or m°H (1 S.D.)
Drift - Less than 1 m°C or m°H per month

Performance Over Operating Conditions Temperature Effects - Less than 1 m°C or m°H per 5°C (9°F) ambient temperature change
Operating Conditions Temperature - 18 to 35°C (64 to 95°F)
Humidity - 5 to 80% relative humidity, (non-condensing)
Storage Temperature - -40 to +45°C (-40 to +113°F)
Electrical Voltage - 100 to 130 VAC (50/60 Hz) or 200 to 250 VAC (50/60 Hz)
Power Consumption 150 Watts
Dimensions 16.0” H x 13.0” W x 18.0” D (40.6 cm x 33.0 cm x 45.7 cm)
Net Weight 28.0 lbs. (12.7 kg)
Compliance ISO5764
  4250
4250 Brochure
  Freezing Point Depression of Authenticated Milk
Freezing Point Depression of Authenticated Milk
 

4250
Click image to enlarge

  Cryoscope User Guide
On Line User Guide
 
         
  4250
MSDS
Freezing Point Calibration Standards
  4250
MSDS
Bath Liquid
 

 

QCL