Gas Consumption For creating standard conditions in a jar suitable for 12 Petri dishes (Mart jar AJ9023), the system consumes gas as follows:
Micro-aerophilic standard recipe: Approximately two litres per jar only
Anaerobic standard recipe: Approximately seven litres per jar only
Jars Per Gas Cylinder Cylinder Micro-aerophilic Anaerobic
10 Litre / 150 bar > 700 jars >200
10 Litre / 200 bar > 950 jars >250
50 Litre / 150 bar > 3,600 jars >1,000
50 Litre / 200 bar > 4,900 jars >1,400
Ultimate Pressure Vacuum Pump 150 mbar
Operating Temperature 10-32°C (50 - 91°F)
Relative Humidity 20-80 %
Voltage 220-240 V
Frequency 50-60 Hz
Peak Current 1.1 Amp
Net Weight 15 Kg
Dimensions W 32 cm, D 55 cm, H 27.5 cm